Urgent Field Safety Notice: Surgical Masks

Yellow triangle warning sign with black exclamation mark

The Bluetree Group has issued an urgent field safety notice relating to Obisk Type IIR Masks.

A problem has been identified with masks manufactured using a particular batch of material. The product meets the requirements for bacterial filtration efficiency, microbial cleanliness and breathability however it does not achieve the required splash resistance to be categorised as a Type IIR. As such it offers infection control equivalent to a Type II surgical mask with the breathability performance of a Type IIR.

Products affected 
Batches of the following products are affected by this Field Safety Notice: 

 Code 

Product Description 

Packaging 

UDI-DI 

Manufacture Date 

OB11R 

Obisk Disposable Type IIR Surgical Masks with Ear Loops 

Supplied in boxes of 50 masks 

05056423700060 

July 2021 

OB11RT 

Obisk Disposable Type IIR Surgical Masks with Tie Backs 

Supplied in boxes of 25 masks 

05056423700220 

July 2021 

 

The following batches of the above products are affected: 

Batch / Lot Number 

Product Code 

101000345 

OB11R 

101000346 

OB11R 

101000348 

OB11R 

101000349 

OB11R 

101000350 

OB11RT 

101000354 

OB11R 

 

More information can be found on the Urgent Field Safety notice relating to Obisk Type IIR Masks provided by The Bluetree Group. 

Please review the content of this field safety notice and check your stock for the affected product code and lot number.

If you have any stock from the affected batch, please isolate them immediately to prevent them from being used or distributed. Any affected batches found can be disposed into standard waste stream.

Published 3rd November 2021